HORIZONTAL AIR FLOW OVENS SMO14HP-2 SMO34HP-2 SMO38HP-2 Previously designated as: HF15-2, HF25-2, HF37-2 INSTALLATION AND OPERATION
10 EXAMPLE PROGRAM Let’s say you want to have the oven to ramp up to 100°C in 30 minutes and control for 1 hour at 100°C. Then you want to ramp
11 at -2 to match the process value to the independent thermometer. To navigate the control to the Offset Calibration mode, push the PGM
12 MAINTENANCE 7.1 Cleaning: The unit chamber should be cleaned and disinfected on a regular basis. Remove shelving & shelving
13 TROUBLESHOOTING TEMPERATURE Temperature too high 1/ controller set too high 2/ controller failed on – call Customer Service 3/ wiring error – ca
14 blower motor in shadow box 3/ is ambient room temperature radically changing – either door opening or room airflow from heaters or air conditionin
15 OTHER Controller on at all times - "locked-up" 1/ Adjust set point to room temperature. If the light goes out but is still heating, r
16 PARTS LIST Description Part No. Blower Motor 4880548 Control Relay 7030533 Convenience Outlet 6100526 Cooling Fan 4880563 Heating Element – SMO14H
17 UNIT SPECIFICATIONS 31.5 x 26.0 x 60.5. Weight Shipping Net SMO14HP-2 (HF15-2) 630 lbs. 445 lbs. SMO34HP-2 (HF25-2) 820 lbs. SMO38HP-2 (HF37-2) 98
18 WIRE DIAGRAM
SHELDON MANUFACTURING, INC. LIMITED WARRANTY (Parts only, exclusive of labor) Sheldon Manufacturing, Inc., (“Manufacturer”) warrants for the origina
2 TABLE OF CONTENTS SECTION 1.0 RECEIVING AND INSPECTION SECTION 2.0 GRAPHIC SYMBOLS SECTION 3.0 INSTALLATION SECTION 4.0 CONTROL OVERVIEW SECTION
3 RECEIVING AND INSPECTION Your satisfaction and safety require a complete understanding of this unit. Read the instructions thoroughly and be sure al
4 GRAPHIC SYMBOLS Your oven has been provided with a display of graphic symbols which is designed to help in identifying the use an
5 INSTALLATION Local city, county or other ordinances may govern the use of this equipment. If you have any questions about local requirements, pleas
6 CONTROL PANEL OVERVIEW 4.1 Power Switch: The main power I/O (on/off) switch controls all power to the unit and must be in the I/ON positi
7 PRECAUTIONS 5.1 The bottom surface of the chamber should not be used as a work surface. 5.2 This unit has been designed with a dampere
8 OPERATION All control functions are covered in the Jumo Manual. This is a simplified version. Unit must be plugged into correct rating
9 1. Push the PGM Button once and then navigate using the Up and Down Arrows until you reach “CONF” in the bottom display. 2. After you reach “CON