HUMIDITY CHAMBER SHC10 SHC10-2 SHC10R SHC10R-2 SHC28 SHC28-2 SHC28R SHC28R-2 PREVIOUSLY DESIGNATED
10 CHART RECORDER INSTALLATION Please note that the following information is a general guide for installation. Before attempting installation pl
11 MAINTENANCE NOTE: Prior to any maintenance or service on this unit, disconnect power cord from the power supply. Service of any electrical compon
12 TROUBLESHOOTING Always make a visual inspection of the unit and control console when troubleshooting. Look for loose or disconnecte
13 1/ verify that controller is asking for heat by looking for Heating light – if pilot light is not on continuously, there is a problem with the co
14 1/ assure that exhaust fan is working 2/ assure that exhaust fan blower casing is free of pooled water 3/ check solid state relay 4/ set point ha
15 Door not sealing 1/ adjust hinge blocks or twist the door. 2/ Confirm that unit has not been damaged and body is not square. Motor doesn't
16 PARTS LISTS SHC10 and SHC10-2 (HC9 and HC9-2) SHC10R and SHC10R-2 (HC9R and HC9R-2) DESCRIPTION SHC10 (HC9) 115V SHC10-2 (HC9-2) 220V SHC10R (HC9R
17 SHC28 (HC30) and SHC28-2 (HC30R) SHC28R (HC30R) and SHC28R-2 (HC30R-2) DESCRIPTION SHC28 (HC30) 115V SHC28-2 (HC30-2) 220V SHC28R (HC30R) 115V SHC
18 UNIT SPECIFICATIONS Weight Shipping Net SHC10 SHC10-2 (HC9 HC9-2) 480 lbs. 341 lbs. SHC10R SHC10R-2 (HC9R HC9R-2) 515 lbs. 401 lbs. SHC28 SHC28-2
RELATIVE HUMIDITY CHART
2 TABLE OF CONTENTS SECTION 1.0 RECEIVING AND INSPECTION SECTION 2.0 GRAPHIC SYMBOLS SECTION 3.0 INSTALLATION AND FACILITIES REQUIREMENTS SECTION
2 WIRING DIAGRAM SHC10 (HC9) (9851449) 1 24FAN1 2412VDCPOWER SUPPLY 6750507OUTINRH SENSOR41005041 234FANBMBLUEWHITEREDBLACKBLOWER MOTOR4880504GREEN L
3 WIRING DIAGRAM SHC10-2 (HC9-2) (9851450) 1 2412VDCPOWER SUPPLY 6750507OUTINRH SENSOR41005041 234FANBMBLUEWHITEREDBLACKBLOWER MOTOR4880504GREEN LIGH
4 WIRING DIAGRAM SHC10R (HC9R) (9851451) 1 24FAN1 2412VDCPOWER SUPPLY 6750507OUTINRH SENSOR41005041 234FANBMBLUEWHITEREDBLACKBLOWER MOTOR4880504GREEN
5 WIRING DIAGRAM SHC10R-2 (HC9R-2) (9851452) 2800503EMI FILTER1 24FAN1 2412VDCPOWER SUPPLY 6750507OUTINRH SENSOR41005041 234FANBMBLUEWHITEREDBLACKBLO
6 WIRING DIAGRAM SHC28 (HC30) (9851453) 1 2412VDCPOWER SUPPLY 6750507OUTINRH SENSOR41005041 234FANBMBLUEWHITERED BLACKBLOWER MOTOR4880504GREEN LIGHTE
7 WIRING DIAGRAM SHC28-2 (HC30-2) (9851454) 1 2412VDCPOWER SUPPLY 6750507OUTINRH SENSOR41005041 234FANBMBLUEWHITEREDBLACKBLOWER MOTOR4880504GREEN LIG
8 WIRING DIAGRAM SHC28R (HC30R) (9851455) 1 24FAN1 2412VDCPOWER SUPPLY 6750507OUTINRH SENSOR41005041 234FANBMBLUEWHITEREDBLACKBLOWER MOTOR4880504GREE
9 WIRING DIAGRAM SHC28R-2 (HC30R-2) (9851456) 2800503EMI FILTER1 24FAN1 2412VDCPOWER SUPPLY 6750507OUTINRH SENSOR41005041 234FANBMBLUEWHITEREDBLACKBL
3 RECEIVING AND INSPECTION Your satisfaction and safety require a complete understanding of this unit, including its proper function and operational
4 GRAPHIC SYMBOLS Your incubator has been provided with a display of graphic symbols which is designed to help in identifying the use and function of
5 INSTALLATION Local city, county or other ordinances may govern the use of this equipment. If you have any questions about local requirements,
6 and assure all volatile or flammable cleaners are evaporated and dry before reattaching the unit to the power supply. 3.6 Humidification Water Sup
7 CONTROL OVERVIEW 4.1 Power Switch: The main power I/O (On/Off) switch controls all power to the unit, and must be in the I/On position before a
8 THEORY OF OPERATION These humidity chambers are designed to maintain temperature and relative humidity at set points controllable by the operator
9 OPERATION It is recommended that your unit be allowed to reach operating temperature before engaging the humidifying system. This requires setti